Description
Xenon arc lamp test chamber reproduce weathering damage by sunlight, temperature, humidity and water spray. Xenon weathering test chambers are utilized in textiles, dyes, leather, plastics, paints, coatings, automotive interior parts, electrotechnical products, color building materials. To perform weathering test, color fastness tests, aging test, hardening test, softening test, crack. Accelerated weathering test methods included ISO4892, ASTM G155-1/155-4, ISO 105-B02/B04/B06, ISO11341, AATCC TM16, TM169, , JIS L0843, SAEJ1960/1885, JASOM346, and many others.
Advantage
● Controlled Irradiance
Controlled Irradiance by narrow band or broad band
● Ensure the stability of the light source
The xenon arc lamp emits high heat when working for a long time. If the lamp is not cooled in time, it will cause the lamp to turn black and affect the life of the lamp. Therefore, LIB has specially designed a set of water cooling system for the lamp, which can continuously provide 18℃ cold water to cool the lamp to ensure continuous operation of at least 1600H.
● Multi-language controller display
Multiple languages can be selected, English, Chinese, Russian, Korean, German, French, Polish, Spanish, Turkish, Romanian
Specification
Model |
XL-S-750 |
Internal Dimension (mm) |
950*950*850 mm |
Overall Dimension (mm) |
1400*1400*1950 mm |
Specimen Holder Size(mm) |
95*200 |
Specimen Capacity |
42pcs |
Chamber Type |
Rotating Holder |
Irradiation Source |
1 piece of 4500w water-cooled xenon arc lamp with inner quartz and outer borosilicate filter |
Irradiance Range |
35~150 W/㎡ |
Bandwidth Measurement |
300nm~400nm (340nm or 420nm) |
Chamber Temperature Range |
Ambient ~ 100 ℃ ±2℃ |
Black Panel Temperature |
BPT 35 ~ 85 ℃ ±2℃ |
Humidity Range |
50% ~ 98% RH |
Humidity Deviation |
± 5% RH |
Water Spray Cycle |
1~9999H59M (Adjustable) |
Controller |
Programmable color LCD touch screen
controller |
Water supply system |
Automatic water supply,Water
purification system |
Air Circulation |
Centrifugal wind fan |
Radiometer |
UV Radiometer, Tolerance: ±5% |
Heating Element |
Nichrome heater |
Cooling System |
Mechanical compression refrigeration
system |
Humidification System |
External isolation, stainless steel surface evaporation
humidifier |
Air Circulation |
Centrifugal wind fan |
Safety Device |
Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth
leakage Protection; Phase Sequence Protection |
Exterior Material |
Steel plate with protective coating |
Interior Material |
|
Thermal Insulation |
Polyurethane foam and insulation cotton |
Observation Window |
Interior lighting, double-layer thermo stability silicone rubber
sealing |
Power Supply |
AC380V 50Hz |
Maximum Noise |
65 dBA |